Etelux Glove Box Regeneration Steps

From WikiName
Jump to navigation Jump to search

1) Regeneration gasoline: a combination of hydrogen and inert gasoline (hydrogen accounts for 5% -10%)

2) Amount: 40L gas cylinder 10Mpa

3) Regenerative fuel stress: .05Mpa (the gas usage per fifty percent hour in the condition of .05Mpa is that the cylinder force drops by 1.5Mpa. If it exceeds one.5Mpa, the force of the force regulator’s schedule will be minimized, if it is much less than one.5Mpa Raise the pressure of the force regulator plan)

four) System: 3 hrs heating-three hours heating and air flow-3 several hours vacuuming-one hour pumping-10 hrs cooling

five) Steps:

① Link the regeneration gasoline (examine the pipeline for leaks right after relationship: immediately after connecting, open up the key valve of the stress lowering valve to the utmost, open the auxiliary meter to .1Mpa, switch off the main valve, wait around 15 minutes, if the strain of the auxiliary meter stays at .one Mpa continues to be unchanged, Acrylic Glove Box indicating no air leakage

>②Set the strain of the auxiliary gauge of the pressure reducing valve to .1Mpa (the pressure for the duration of ventilation is about .06-.08MPa

>③ Near the cycle and analyz

r>④Turn on the vacuum pu

r>⑤ Begin regenerat

br>⑥ The new application ends at 20h (thoroughly clean in advance of cleaning, h2o and oxygen index <200p


6) Note: Do not cut off the power and gas during the regeneration process (especially the first 6 hours)! Some pressure reducing valves are not allowed, and the pressure needs to be adjusted appropriately to ensure continuous ventilation for 3 ho


7) Observation: whether the regeneration gas flowmeter shows about 20L / min, otherwise adjust the pressure of the pressure reducing valve.